Abstract
BACKGROUND AND PURPOSE: Although digital subtraction angiography (DSA) provides excellent visualization of the intracranial vasculature, it has several limitations. Our purpose was to evaluate the ability of helical CT angiography (CTA) to help detect and quantify intracranial stenosis and occlusion compared with DSA and MR angiography (MRA).
METHODS: Twenty-eight patients underwent CTA, DSA, and 3D time-of-flight (TOF) MRA for suspected cerebrovascular lesions. All three studies were performed within a 30-day period. Two readers blinded to prior estimated or calculated stenoses, patient history and clinical information examined 672 vessel segments. Lesions were categorized as normal (0–9%), mild (10–29%), moderate (30–69%), severe (70–99%), or occluded (no flow detected). DSA was the reference standard. Unblinded consensus readings were obtained for all discrepancies.
RESULTS: A total of 115 diseased vessel segments were identified. After consensus interpretation, CTA revealed higher sensitivity than that of MRA for intracranial stenosis (98% versus 70%, P < .001) and occlusion (100% versus 87%, P = .02). CTA had a higher positive predictive value than that of MRA for both stenosis (93% versus 65%, P < .001) and occlusion (100% versus 59%, P < .001). CTA had a high interoperator reliability. In 6 of 28 patients (21%), all 6 with low-flow states in the posterior circulation, CTA was superior to DSA in detection of vessel patency.
CONCLUSION: CTA has a higher sensitivity and positive predictive value than MRA and is recommended over TOF MRA for detection of intracranial stenosis and occlusion. CTA has a high interoperator reliability. CTA is superior to DSA in the evaluation of posterior circulation steno-occlusive disease when slow flow is present. CTA results had a significant effect on patient clinical management.
Stroke is the third leading cause of mortality and the leading cause of morbidity in the United States. Atherosclerotic disease results when soft and calcific plaque accumulates in the vessel wall leading to luminal narrowing. The presence of plaque predisposes to vascular thrombosis or ischemic states leading to stroke (1). Lutsep and Clark (2) found that 15% of patients with cortical symptoms or signs of stroke have intracranial stenosis of 50% or greater at MR angiography (MRA) or conventional angiography. Studies have shown the potential benefit of angioplasty, with or without stent placement, of symptomatic stenotic intracranial arteries (3, 4). Given recent advances in pharmacologic treatment for stroke risk factors, including atherosclerosis, and an emerging literature on endovascular therapies for intracranial stenosis, it is important to both detect and accurately assess the severity of intracranial stenoses.
Digital subtraction angiography (DSA) has long been considered the reference standard for imaging evaluation of intracranial stenosis and occlusion. DSA provides excellent visualization of the intracranial vasculature. However, it has several limitations including a high skill required to perform the study and the necessity for intraarterial catheterization, which results in increased morbidity and mortality for the patient, with a 0.7% stroke risk associated with each procedure (5).
Given the disadvantages of DSA, CT angiography (CTA) and 3D time-of-flight (TOF) MRA have emerged as feasible noninvasive techniques to evaluate the intracranial vasculature. The purpose of this study was to evaluate the ability of CTA to help detect and quantify intracranial stenosis and occlusion compared with DSA and 3D-TOF MRA.
Methods
We retrospectively compared CTA, DSA, and MRA images of 28 patients who underwent these examinations between May 1997 and November 2000 for suspected intracranial atherosclerotic disease. All three studies were obtained within 30 days of each other to reduce the possibility of dynamic vascular change affecting vessel patency over time. Inclusion criteria were symptoms of cerebral ischemia with or without neurologic deficit and transient ischemic attack. Exclusion criteria were the CTA, DSA, and MRA examinations separated by more than 30 days, image quality inadequate for interpretation, known or suspected vasospasm, and interposed intraarterial or intravenous thrombolysis or clot retrieval therapies. In one patient, technical limitations rendered the CTA images inadequate for interpretation in the basilar artery, and this vessel was excluded from evaluation in all modalities. Institutional review board approval for the study design was obtained.
Image Acquisition
Standard clinical protocols were used for CTA, DSA, and MRA with an emphasis on high spatial resolution.
The CTA protocol involved a 120-kV peak, 280–300 mA, 18-cm FOV, 512 × 512 matrix, 1.0–1.5 pitch, 1-mm section collimation, 0.35 × 0.35 × 0.5-mm voxel size, 0.5–0.625-mm reconstruction interval, 9–13-cm slab thickness, and 30–60-second scanning time. Images were acquired with a single- (CTi 9800; GE Medical Systems, Milwaukee, WI) or multi–detector row (Lightspeed 4 Detector; GE Medical Systems) scanner. Patients received 120–200 mL of intravenous iohexol (Omnipaque 350; Amersham Health, Princeton, NJ) delivered through a 22-gauge or larger antecubital catheter at 3 mL/s. A timing run was performed before every CTA study. Timing runs were obtained at the level of the midneck. Contrast material 20 mL was delivered at 3 mL/s, and 15 images are obtained at the same level by using a 1-second scan with a 1-second interscan delay following a 10-second initial delay. A region-of-interest graph plotting attenuation (in HU) versus time was then generated at the console to determine the optimal time delay for the helical scan. Total delay was obtained by adding the up-front delay to the timing delay.
The DSA protocol involved femoral puncture and selective injection of cervical vessels employing the Seldinger technique, by using 6–9 mL of iohexol (Omnipaque 240) per injection. The DSA runs were obtained at 3–4 frames per second to late venous phase, by using a matrix of 1024 × 1024, 15-cm FOV, pixel size 0.15 × 0.15. Anteroposterior, lateral, and oblique projections were acquired in all cases. Magnification views were acquired when needed to clarify significant findings.
The MRA protocol was a 3D TOF MR technique performed with a 1.5-T Magnetom Vision (Siemens, Erlangen, Germany) or LX (GE Medical Systems) unit with use of the following parameters: 35/7.2 (TR/TE), 200 × 512 matrix, and 20-cm FOV. Postprocessing was performed by using a maximum intensity projection (MIP) rendering, generating 22 total, 16.4° incremental rotations in somersault and clockwise rotation. Superior saturation pulse was employed. Source images were generated.
CTA postprocessing was performed on a Vitrea workstation, versions 3.0–3.3 (Dual Processor 850 MHz-1GHz Pentium III) by using perspective volume rendering for lesion detection. Two-dimensional, gray-scale, axial images were used in all cases for lesion quantitation. Two-dimensional, gray-scale, sagittal and coronal multiplanar reformations (MPRs) and curved oblique reformations were generated as necessary in all cases of suspected disease for lesion confirmation and characterization.
Image Analysis
Part I: Comparative analysis of CTA, DSA and MRA. The CTA images were reviewed prospectively for presence, location, and severity of lesions by a single independent blinded reader and retrospectively for lesion presence and severity by a second blinded reader. Both blinded readers (S.B., P.V.) prospectively reviewed DSA and MRA images for lesion presence, location, and severity.
Each imaging technique data set was processed completely, and the cases were presented in a random order. Both readers were blinded to all prior estimated or calculated stenoses as well as to the patient’s history and all clinical information. Both radiologists independently characterized each stenotic lesion as focal or segmental, and eccentric or concentric.
All CTA, DSA, and MRA lesions were measured by using visual estimation of the full-width half-maximum of the section sensitive profile at the margins of the opacified vessel.
For all imaging modalities, the North American Symptomatic Carotid Endarterectomy Trial (NASCET) criteria were used for stenosis calculations: [(Dn− Ds)/Dn] × 100, where Dn is normal diameter and Ds is stenosed diameter. NASCET stenoses were grouped according to the following grading scale: normal (0–9%), mild (10–29%), moderate (30–69%), severe (70–99%), or occluded (no flow detected).
Diseased vessels were defined as those whose NASCET stenosis rates fell into the moderate, severe, or occluded categories. Specifically, stenoses graded as normal (0–9%) and mild (10–29%) were not considered diseased vessel segments and were excluded from analysis.
On DSA images, a lesion was defined as occluded if no flow was seen in a portion of a vessel segment either with anterograde injection or via reflux. On MRA images, a lesion was defined as occluded if there was a long-segment flow void. Short-segment flow voids were assumed to represent severe stenoses. On CTA images, an occlusion was defined as nonvisualization of a vessel segment on the 2D MPR images.
CTA stenoses were detected by using the 3D image data, with free-form rotation of the arterial volume centered at all major branch points of the circle of Willis, and by using 22 standard image projections in both clockwise and tumble rotation directions. After identification of a candidate stenosis, reduction in luminal caliber was documented on the 2D gray-scale MPR images. Lesion measurements were obtained from the 2D MPR images, always orthogonal to the longitudinal axis of the artery. CTA stenosis quantitation was performed using the 2D gray-scale MPR images with approximate window and level settings of 1000 and 500 HU, respectively. Digital internal calipers were used for CTA lesion measurements with linear accuracy of ± 10% for measurements less than 2 mm and ± 4% for measurements of 2–10 mm.
DSA stenosis quantitation was performed with Mitutoyo digital calipers (Mitutoyo Corporation, Japan). Single frame, middle arterial phase angiogram views were used for DSA measurements. To account for magnification differences, all DSA measurements were obtained from equal magnification views, with the assumption based on published reports of pathologic specimens that the average adult supraclinoid internal carotid artery (ICA) measures 3.63 mm (6). Proportionality equations were then generated to calculate normal diameter (Dn), stenosed diameter (Ds), and percentage stenosis.
MRA stenosis quantitation was performed with Mitutoyo Digimatic digital caliper. MRA measurements were obtained on the MIP image showing the narrowest luminal diameter and the first normal distal diameter with parallel walls. Both the clockwise rotation and tumble projection images were reviewed. Digital caliper apertures were scaled to the sidebar image scale to obtain the actual stenotic and normal luminal dimensions. Source images also were used to confirm stenosis location and severity.
Assessed intracranial vessels included the following bilateral arterial segments: high cervical ICA, petrous ICA, cavernous ICA, supraclinoid ICA, A1, A2, M1, M2, intracranial vertebral, P1, and P2, as well as the proximal basilar and distal basilar artery. Therefore, a total of 672 vessel segments were examined in the 28 patients. These vessel segments were chosen because they are large and medium-sized arterial segments generally visualized with all three imaging modalities (DSA, CTA, and MRA). The anterior inferior cerebellar artery, posterior inferior cerebellar artery, anterior communicating artery, and posterior communicating artery were not assessed, as these vessels often measure less than 1 mm and were not consistently seen with all three imaging modalities. If an intracranial artery was uniformly and segmentally of decreased caliber throughout its length, compared with the contralateral side, this vessel segment was considered hypoplastic, and no stenosis was attributed.
Standard contingency tables were generated for sensitivity, specificity, and positive and negative predictive values by using DSA stenosis and occlusion values as the reference standard.
Part II: Interoperator reliability. The interoperator reproducibility for the severity of CTA lesion grading was assessed. Two separate readers independently measured lesion severity on CTA images in a blinded fashion. NASCET criteria were reapplied by the second blinded reader to the preidentified and graded sites of stenosis by the initial interpreting neuroradiologist.
Data Analysis
Part I: Comparative analysis of CTA, DSA, and MRA for intramodality discordancy. After the blinded interpretation of all modalities by both independent readers, the DSA, CTA, and MRA images were then reviewed by both readers to obtain consensus in cases of intramodality discordant readings, where the presence of a stenosis was in question. During intramodality discordancy evaluation, the readers remained blinded to the interpretation of the other two modalities.
Intramodality discordant readings were defined as those occurring when one reader made an interpretation error (“reader error”). This was recorded on the Reader Error Rate Chart (Table 2). The intramodality discordant cases, where the presence or absence of a stenosis or occlusion within a vessel segment of a specific technique was not initially in agreement between both readers, were included in the contingency tables for sensitivity, specificity, positive and negative predictive values, and accuracy, only after a consensus decision regarding disease state had been agreed on.
Part II: Comparative analysis of CTA, DSA, and MRA for intermodality discordancy. The DSA, CTA, and MRA images were then reviewed in a side-by-side fashion by both readers to obtain consensus of true disease status in cases of intermodality discordant readings. For example, if DSA demonstrated an occlusion in a vessel segment, but the correlative CTA image demonstrated a patent but stenotic lumen, the lesion was either interpreted as a probable false-positive finding for occlusion at DSA or probable interval restoration of blood flow. This conclusion was based on the assumption that a contrast material-opacified vessel reflects true vessel patency. Probable interval spontaneous restoration of blood flow was concluded if there was no significant stenosis in the vessel proximal to the apparent occlusion. In this manner, false-negative findings for vessel occlusion were also assessed. The P values were calculated by using Bennett’s χ2 statistic (7).
Part III: Interobserver reliability. To assess interobserver reliability for CTA, the calculated stenosis measurements of the first reviewer were compared with those of the second reviewer. Vessel caliber was measured to grade lesions from 0% (no stenosis) to 100% (occluded). A Pearson correlation coefficient value and mean reader difference values were calculated to assess interobserver reliability in lesion grading.
Results
Of the 672 intracranial arterial segments examined, 115 diseased artery segments (17%) were identified with DSA. Again, diseased vessel segments were defined as those segments that fell into the moderate or severely stenotic, or occluded categories. Only a small number of vessel segments fell into the mildly stenosed category (10–29%), and these vessels were not characterized as diseased because they were considered unlikely to be of clinical relevance. Therefore, the mildly stenosed vessel segments were not used in the contingency tables to assess the sensitivity, specificity, positive predictive value, and negative predictive value of each technique.
There was a heterogeneous distribution of disease throughout all vessel segments examined (Table 1), with anterior circulation vessels composing 42% and posterior circulation vessels composing 58% of all lesions encountered.
The female-to-male ratio was 12:16 (43:57%). Age range was 24–79 years, with a mean age of 58 years. The mean, median, and mode time of separation of the three studies was 1 day, with a range of 0–30 days.
Each patient had a variable number of diseased vessel segments with a mean of 4, mode of 3, and range of 1–13 per patient. The number of mildly stenotic vessel segments identified with each technique in our patient population was low (DSA, n = 5; CTA, n = 3; and MRA, n = 1). Also, the number was low for lesions classified as severely stenotic, and a proportionately greater number of severely stenotic lesions were found with CTA and MRA compared with DSA (DSA, n = 7; CTA, n = 15; and MRA, n = 16). Most intracranial lesions were of moderate severity, with an equal distribution across modalities.
Interoperator reliability was high, with a Pearson correlation coefficient (r) of 0.951 for grading of lesion severity with CTA. The 95% confidence interval (CI) for the Pearson correlation coefficient was 0.93–0.96. The mean difference between the two readers was 0.01611 with a standard deviation of 0.09794 and 95% CI of −0.002 to 0.034.
The reader error rate for each lesion by each technique as defined by consensus decision by both readers was relatively low across all imaging modalities (Table 2). Most reader errors across all imaging modalities occurred when the stenoses were eccentric, and the optimal projection or viewing angle demonstrating the minimal diameter for stenotic segments was not initially recognized.
No procedural complications occurred for any technique. Average setup and image acquisition time and processing times for CTA were 12 and 22 minutes, respectively. Similar times for 3D-TOF MRA were 7 and 7 minutes, respectively. For DSA, comparable times were 15 and 10 minutes, respectively. No complete technical failures occurred for any technique, with five CTA, eight MRA, and two DSA studies showing some patient motion or other artifacts, which precluded image analysis of one vessel in one patient.
Of the 115 diseased vessel segments evaluated, we found seven vessel segments (in seven patients) that demonstrated intermodality discrepancies in lesion classification as occlusion or stenosis. In one of these patients, CTA was normal, but DSA demonstrated segmental occlusion. In this patient, spontaneous lysis of the thrombus probably occurred in the 72 hours between the time DSA and CTA were performed, because image quality was excellent in both studies and no technical issues were identified. Also, the patient improved clinically between the two studies.
Of the remaining six patients, six suspected reference standard (DSA) errors were identified. Errors were defined as errors in the depiction of true vascular anatomy made by DSA following consensus reevaluation of all the data. These intermodality discrepancies were documented as DSA false-positive findings (Table 3). A DSA false-positive finding was defined as DSA demonstrating occlusion of a vascular segment by consensus visual inspection, whereas CTA (and in two cases MRA) demonstrated a patent but stenotic lumen. All of the DSA false-positive findings occurred in the posterior circulation. Of the 115 lesions identified with DSA, six vessel segments in six patients erroneously appeared occluded on DSA images, indicating probable DSA false-positive results in six (21%) of 28 patients. The suspected DSA false-positive findings occurred in six (23%) of 26 of all visualized intracranial occlusions, in six (9%) of 67 posterior circulation lesions, and in six (5%) of 115 intracranial lesions identified with DSA.
We observed that the probable DSA false-positive findings occurred in very slow- and/or low-flow conditions present distal to a severe stenosis in a posterior circulation artery, causing the vessel to appear occluded at DSA, but stenotic at CTA (Figs 1–3). By using DSA as the reference standard, these DSA false-positive findings artificially elevated the false-negative fraction for CTA occlusion data, lowering the sensitivity (Table 4). Likewise, the DSA false-positive findings artificially elevated the false-positive fraction for the CTA stenosis data, lowering the positive predictive value (Table 5).
In the comparative analysis chart for intracranial occlusions (Table 4), the first percentages represent our results using DSA as the reference standard. The percentages in parentheses are corrected for the probable DSA false-positive findings by consensus reading. The lowered sensitivity of CTA for intracranial occlusions when DSA is used as the reference standard becomes 100% when the proposed DSA interpretation error is removed. Of note, the positive predictive value remains low for MRA secondary to the high false-positive fraction of this technique.
The comparative analysis chart for intracranial stenosis (Table 5) demonstrates the artificially lowered positive predictive value for CTA, which adjusts to 93% when the proposed DSA interpretation error is removed. Again, note the fixed low positive predictive value for MRA secondary to the high fixed false-positive fraction of this technique. Also, note the comparatively lower sensitivity of MRA than CTA for detection of stenotic lesions.
The raw numbers used to construct the comparative analysis charts are seen in Tables 6–9. The Tables are corrected for reader error as well as DSA false-positive findings to reflect the true state of disease. CTA had a higher sensitivity than MRA for intracranial stenosis (98% versus 70%, P < .001) and occlusion (100% versus 87%, P = .02). CTA had a higher positive predictive value than MRA for both stenosis (93% versus 65%, P < .001) and occlusion (100% versus 59%, P < .001), reflecting a comparatively lower false-positive rate for CTA. The negative predictive value of both CTA and MRA was high for detection of both stenosis (100% versus 98%, P < .001) and occlusion (100% versus 99.5%, P < .001).
Discussion
DSA provides excellent visualization of the intracranial vasculature and has been considered the reference standard for evaluation of intracranial stenosis and occlusion. This is due to the high spatial resolution that this technique generates with a very small pixel size. Dynamic flow information of the cerebrovascular arterial tree is provided with each injection. However, given the risks, costs, and limited availability of conventional angiography, the need for noninvasive techniques is growing.
TOF MRA is a widely available noninvasive technique that requires no radiation exposure and no administration of contrast material. Limitations of TOF MRA include a lower spatial resolution, as compared with DSA and CTA, that may diminish the ability to depict severe stenosis in small vessels and a dependence of vascular flow signal intensity on the maintenance of proton spin coherence. Stenotic lesions are sites where proton spin dephasing, and consequently flow signal intensity loss, commonly occur. This is due to complex, slow or in-phase flow, all flow characteristics commonly seen in stenotic intracranial vessels (8). In addition, the TOF technique requires comparatively long imaging times, frequently leading to patient motion artifacts and degradation of image quality. Although the literature demonstrates sensitivities in the range of 80% for accurately categorizing stenotic or occlusive lesions with MRA (8, 9), these limitations create ambiguity. We occasionally found it very difficult to determine whether we were seeing flow artifact, true stenosis, or even occlusion in a vessel. If flow signal intensity decreases caused by disturbed flow dynamics are sufficient to create a flow gap on a TOF image, this may necessitate performing additional, possibly more invasive techniques such as DSA. Despite these limitations, TOF MRA with MIP requires minimal user interaction to acquire and process, and it uses a facile display mode. Our MRA results demonstrated a 70% sensitivity for detection of intracranial stenosis and an 81% sensitivity for detection of intracranial occlusion when compared with the reference standard of DSA, values comparable to those reported in the literature (8, 9). Our positive predictive value for the quantitation of intracranial stenosis by using TOF MRA (65%) was equal to that recently reported in the Stroke Outcomes and Neuroimaging of Intracranial Atherosclerosis study, which found a positive predictive value of 64% (10). Our current MRA postprocessing does not use targeted MIPs. Possibly our MRA results could have been improved through the use of such techniques. However, we did not experience instances where overlapping vessels precluded lesion detection or quantitation.
Helical CTA is a noninvasive technique that provides good visualization of the intracranial vasculature (11). Our routine CTA protocol requires 60 seconds or less to perform and has higher spatial resolution relative to that of MRA, but lower than that of DSA. Limitations of CTA include radiation exposure as well as time and skill involved in image postprocessing. Other potential limitations to CTA include interpolation errors, contrast material entry phenomenon, possible venous contamination in a region of interest, and the need for optimized contrast gradient-injection timing. Because CTA is essentially a vessel cast technique, it also provides no significant flow hemodynamic information.
We have not found venous contamination to be a significant problem when evaluating data sets. The veins are normally opacified in CTA studies performed on single detector helical scanners, since scanning time is approximately 30–60 seconds. Even if a vein overlaps an artery of interest in one or more planes, the arterial segment is usually accurately identified by its smooth contour, higher HU attenuation, and characteristic spatial location.
Skutta et al (12) suggested that CTA may be superior to MRA in depicting poststenotic or postocclusive arterial segments of low flow or turbulence. Our data agrees with this finding and further emphasizes the limitations of routine TOF MRA. Gadolinium-enhanced MRA of the intracranial vasculature, not currently widely available because of technical issues, may alleviate some of the limitations of TOF MRA by decreased dependence of flow signal intensity on TOF entry phenomenon and decreased T1 shortening effects of fat and calcium, but spatial resolution issues remain.
A recent study compared the combined results of both CTA and MRA with those of the current reference standard, DSA, for evaluation of intracranial stenosis (13). In that study, the CTA images were not blindly assessed for lesion presence, as the CTA lesion location was determined according to the suspected location of stenosis on the previously read MRA images. Those authors found that the combined accuracy of MRA and CTA is equal to that of DSA in most cases. Results of another study indicate that CTA had difficulty in enabling identification of stenotic lesions in the vertebral artery, but otherwise suggest that CTA is a reliable method for detection of lesions in the posterior fossa and may replace DSA in many cases (14).
After initial blinded image analysis, the DSA, CTA, and MRA images in our study were reviewed in a side-by-side fashion by both readers to obtain consensus in cases of intermodality discordant readings. This approach enabled us to identify an imaging technique error independent of “reader error.” We encountered six instances in which DSA falsely demonstrated arterial occlusion in the posterior circulation, indicating a 23% rate of DSA false-positive findings for arterial occlusion, occurring in 21% of our patient population. MRA also had false-positive findings for occlusion in four of the six cases. Examination of the angiographic runs from these patients revealed that all of the DSA false-positive instances occurred in low- or balanced flow states in the setting of a severe stenosis in the posterior circulation (Figs 1 and 2) or in the setting of a relatively isolated posterior circulation (Fig 3). Four of the DSA false-positive findings occurred in the basilar artery and two in the vertebral artery.
We hypothesize the reason the vessel distal to the stenosis appeared occluded on DSA images, when CTA clearly showed the vessel to be patent, is related to differences in image acquisition time for DSA and CTA modalities. Angiographic runs are obtained during the course of a single intracranial circulation cycle, typically 5–7 seconds on average, by using a 3–4 frame per second filming rate, whereas CTA studies are acquired over approximately 60 seconds by using a single detector and 30 seconds by using a four detector helical scanner, allowing the contrast material to circulate up to 4–12 times in the intracranial vessels before imaging, which permits flow of a larger volume of contrast material through a tight stenosis. This greater volume of contrast material distal to a tight stenosis coupled with the HU-based methods of transmission tomography, may help to make these vessel segments more conspicuous, thus rendering visibly stenotic, but patent, arterial segments on CTA images that appear occluded on DSA images. This comparative advantage of CTA may be lost with the newer generation, faster multi-detector row scanners, where routine protocol total scanning times may approach conventional angiographic exposure times. Therefore, for instances in which a vertebrobasilar occlusion is suspected, we recommend a lower rate of table translation per unit time and/or delayed imaging when patients are scanned with multi-detector row systems.
The ability to correctly identify a patent vessel has important clinical and therapeutic implications. If a vessel is truly occluded, it may not be necessary to initiate aggressive antithrombotic therapy, and the probability of successful endovascular therapeutic intervention is lower. However, if the vessel is in fact tightly stenotic but patent, a significant risk of thromboembolic stroke may exist, thus justifying aggressive antithrombotic therapy. In addition, endovascular intervention, such as angioplasty or stent placement, and neurosurgical options, such as arterial bypass procedures, can be considered.
This study has several limitations and will benefit from confirmation with studies using larger samples. One limitation of our study is that our time limit for obtaining all three studies was 30 days. It is possible that a patient’s vessel may have been occluded at the time of DSA, but only stenosed at the time CTA was performed. The mean, median, and mode time difference of our three imaging modalities was 1 day, but inclusion criteria permitted inclusion of patients with studies separated by a period of up to 1 month. The mean time difference in imaging studies for the six patients in which DSA false-positive findings occurred was 72 hours, and no changes in neurologic status were observed during the intervening time period in any of these patients to suggest a major change in circulation patterns had occurred. A systematic bias could have been introduced if one technique routinely preceded the other. However, we evaluated our data and did not find such a pattern.
Another limitation is that the comparison between the three modalities was retrospective, and inclusion criteria for the study were not prospectively established, possibly introducing a selection bias. This means that the frequency of DSA false-positive findings in this study population cannot be generalized to all patients with symptomatic intracranial stenosis, since the criteria for ordering the imaging studies were not known for each case. The use of consensus readings in this study further introduces a possible reader bias. Another potential limitation was possible introduction of a systematic error through the use of MIP images for quantitation when MRA was being evaluated, but source oblique MPR images for CTA quantitation. This approach was taken because the source images for the MRA data sets were thought to provide vessel boundaries that were not easily measured, leading to decreased confidence in vessel boundary determinations, and therefore lower certainty in stenosis determinations. Finally, both CTA and MRA are techniques in evolution, and further protocol refinements could significantly alter study results. It is possible that interpretation of the CTA data by using volume-rendering techniques introduced a systematic error due to inherent technical advantages or limitations of volume rendering over MIP techniques used with MRA. We found volume rendering of the MRA data was not a feasible approach owing to the lower spatial resolution and anisotropic nature of routine clinical TOF MRA sequences.
Previous articles have reported that CTA is limited in its ability to evaluate the petrous ICA (12), cavernous ICA (13), and vertebral artery (14). We did not encounter this problem in any of our patients. We suggest that prior difficulty may be eliminated by close attention to data acquisition and processing. This requires ensuring optimal contrast material opacification within the vessels through monitored timing injections or automated bolus detection software. It is our opinion that this results in better data sets than when no timing run is used, as others have reported significant variations in peak contrast material opacification across individuals (15). Other recommendations for increasing conspicuity of the vessels are a fast injection rate (3–4 mL/s), high contrast material concentration (350–370 mmol), and a small FOV to decrease voxel size and increase spatial resolution. Finally, all intraosseus vessel segment stenoses and occlusions must be detected by using multiplanar or curved multiplanar 2D gray-scale images, by using appropriate window and level settings, rather than the 3D images, since the latter may not be able to discriminate between pixels belonging to bone and those belonging to arteries, due to the overlapping HU attenuation profiles of both structures.
The choice of postprocessing software and the manner in which postprocessing is performed is critically important for accurate CTA interpretation. This is likely the most important factor accounting for the variable success of vessel characterization in the CTA literature. We favor using perspective volume rendering to view 3D data because this technique makes available for interrogation all the voxels within a volume, while avoiding potential arbitrary vessel border definition and extensive loss of information, a limitation inherent to shaded surface display and MIP techniques (16, 17). Volume rendering is now the preferred rendering technique for CTA (18). Others have also found that perspective volume rendering is superior to MIP (19) and shaded surface display for evaluating vessels (20). Perspective volume rendering is the only postprocessing technique that allows visualization of both nonluminal and luminal vascular information, including a clear distinction between soft and calcific plaque versus contrast-opacified vessel. Detection of noncalcified plaque is important because studies have shown that aggressively treated soft plaque may regress under maximal medical management (21). We have found that 2D source MPR images or curved oblique MPR images are most optimal for evaluating vessels surrounded by bone and the 3D volumetric windows are most optimal for evaluating vessels surrounded by brain. The reader can dynamically interact with the data set and freely adjust window and level settings to maximize vessel conspicuity in vessels bordered by bone structures, such as the petrocavernous segments of the ICA.
The learning curve required to postprocess CTA data is a limitation of the technique. If the window and level settings are not correct, a stenosis can be overestimated or underestimated (blooming effect). Failure to optimally adjust window and level settings may also result in inability to visualize calcific and soft plaque or contrast-opacified lumen and bone (despite the quite different HU). The correct window and level settings are defined as the settings that will optimize conspicuity of the vessel without resulting in significant blooming or margin obscuration by adjacent bone or soft tissues. For this study, 3D image review was conducted with values set at a window of approximately 250 HU and level of 150 HU, whereas for the 2D image data the window and level settings were approximately 1000 and 500 HU, respectively.
Our goal is that this analysis may help clarify the relative value of each technique in the evaluation of patients with intracranial steno-occlusive disease. In our study, each of the six patients in whom CTA demonstrated a posterior circulation artery to be patent and not occluded, as suggested by DSA, received aggressive medical therapy with Coumadin and/or Lovenox. In addition, one patient underwent stent placement. This indicates that the CTA results had a significant effect on patient clinical management.
We suggest CTA be considered as the primary study in the setting of suspected posterior circulation occlusion. If CTA demonstrates that the artery is patent, then DSA can be considered for evaluation of flow hemodynamics and possible intervention.
Conclusion
CTA demonstrates a high sensitivity, specificity, and positive predictive value for the evaluation of intracranial stenosis and occlusion. CTA has higher sensitivity and positive predictive value than those of MRA and is recommended over TOF MRA for the evaluation of stenotic and occluded intracranial vessel segments. CTA has a high interoperator reliability for quantitation of stenotic lesions when expert readers are used. Helical CTA is superior to DSA in the demonstration of arterial patency in posterior circulation arteries when very low- or balanced-flow states are present due to a severe stenosis. CTA results had a significant effect on patient clinical management.