Abstract
BACKGROUND AND PURPOSE: Quantitative MR imaging techniques are gaining interest as methods of reducing acquisition times while additionally providing robust measurements. This study aimed to implement a synthetic MR imaging method on a new scanner type and to compare its diagnostic accuracy and volumetry with conventional MR imaging in patients with MS and controls.
MATERIALS AND METHODS: Twenty patients with MS and 20 healthy controls were enrolled after ethics approval and written informed consent. Synthetic MR imaging was implemented on a Siemens 3T scanner. Comparable conventional and synthetic proton-density–, T1-, and T2-weighted, and FLAIR images were acquired. Diagnostic accuracy, lesion detection, and artifacts were assessed by blinded neuroradiologic evaluation, and contrast-to-noise ratios, by manual tracing. Volumetry was performed with synthetic MR imaging, FreeSurfer, FMRIB Software Library, and Statistical Parametric Mapping. Repeatability was quantified by using the coefficient of variance.
RESULTS: Synthetic proton-density–, T1-, and T2-weighted images were of sufficient or good quality and were acquired in 7% less time than with conventional MR imaging. Synthetic FLAIR images were degraded by artifacts. Lesion counts and volumes were higher in synthetic MR imaging due to differences in the contrast of dirty-appearing WM but did not affect the radiologic diagnostic classification or lesion topography (P = .50–.77). Synthetic MR imaging provided segmentations with the shortest processing time (16 seconds) and the lowest repeatability error for brain volume (0.14%), intracranial volume (0.12%), brain parenchymal fraction (0.14%), and GM fraction (0.56%).
CONCLUSIONS: Synthetic MR imaging can be an alternative to conventional MR imaging for generating diagnostic proton-density–, T1-, and T2-weighted images in patients with MS and controls while additionally delivering fast and robust volumetric measurements suitable for MS studies.
ABBREVIATIONS:
- BPF
- brain parenchymal fraction
- BV
- brain volume
- CoV
- coefficient of variance
- GMF
- gray matter fraction
- ICV
- intracranial volume
- PD
- proton density
- SPM
- statistical parametric mapping
- SyMRI
- synthetic MR imaging
- WMF
- white matter fraction
In conventional MR imaging, multiple sequences with different contrast weightings are obtained. This process is time-consuming with redundant data acquisition. Techniques such as MR fingerprinting and synthetic MR imaging can reduce acquisition times and thereby increase MR imaging availability for both clinical applications and research.1⇓–3 SyMRI is a synthetic MR imaging method based on a quantitative approach in which a single saturation recovery TSE sequence is used to estimate absolute physical properties, the proton density (PD), longitudinal relaxation rate, and transverse relaxation rate, including correction for B1-inhomogeneities. Rather than predetermining acquisition parameters such as TE, TI, and TR to maximize tissue contrast,3 synthetic MR imaging produces a free range of synthetic weightings based on a single sequence through mathematic inference.4,5 The quantitative nature of the method and its ability to probe multiple physical properties in a single sequence make it suitable for volumetric analysis.6⇓⇓⇓–10 Synthetic MR imaging has shown promising initial results for use in MS and patients with an ischemic event.11,12 The technique is consequently gaining interest as a potentially time-efficient alternative to conventional MR imaging to visualize and quantify brain tissue properties.
MS is a chronic neuroinflammatory disorder affecting 2.5 million people globally.13 MS has a heterogeneous clinical expression, which complicates the choice of disease-modifying therapy.14 MR imaging is a cornerstone for the diagnosis and monitoring of MS, but qualitative MR imaging measurements are poorly correlated with the clinical outcome. Volumetric MR imaging measurements have an independent predictive value in MS but require laborious image postprocessing, limiting its clinical potential.15,16 Robust and fully automatic volumetry approved for clinical use would thus be important for clinical care and research purposes. The synthetic MR imaging technique has initially been developed for use on Philips (Best, the Netherlands) and GE Healthcare (Milwaukee, Wisconsin) MR imaging systems, but it is not available for other systems and independent evaluations of the method are scarce.5,6
The purpose of this study was to implement the synthetic MR imaging technique for use on Siemens (Erlangen, Germany) MR imaging scanners and to compare the diagnostic accuracy of synthetic and conventional images in MS. A secondary aim was to test the repeatability of the volumetric synthetic MR imaging measurements and compare the volumetric results and practicality with other commonly used brain volumetric methods.
Materials and Methods
Participants and Clinical Assessment
In this prospective study, 20 patients with MS were recruited from the MS center at Karolinska University Hospital. A senior consultant in neurology (S.F.) performed neurologic examinations and rated physical disability by using the Expanded Disability Status Scale, blinded to the radiologic evaluations. Characteristics for patients with MS were the following: 16 women, 4 men; mean age, 44 ± 14 years. MS subtypes were 9 relapsing-remitting, 10 secondary-progressive, and 1 primary-progressive. The mean disease duration was 17.4 ± 11 years; median Expanded Disability Status Scale score, 2.5 (range, 1.0–8.5) with disease-modifying therapy in 12 patients (60%). A group of 20 age-matched neurologically healthy controls (11 women, 9 men; mean age, 41 ± 17 years) was also recruited. The ethics review board in Stockholm approved the study, and written informed consent was obtained from all participants.
Image Acquisition
Technical details for all MR imaging sequences are given in Table 1. Synthetic MR imaging was implemented as a saturation recovery TSE sequence with a saturation pulse flip angle of 120°, a multiecho readout including phase and magnitude data, with 4 repetitions on a Magnetom Trio 3T scanner (Siemens) on the VB17 software platform using a 12-channel head coil. The section acquisition order was altered for each repetition, resulting in 4 different effective saturation delays (150, 580, 2000, and 4130 ms) for each section. The resulting data formed a matrix of 8 complex images per section at different saturation delays and TEs. A least squares fit was performed on the signal intensity of these images as previously described,3 resulting in estimates of the longitudinal and transverse relaxation rates, PD, and B1 field inhomogeneities. The synthetic MR imaging acquisition time was 6:50 minutes. Conventional PD-, T1-, and T2-weighted, and FLAIR images were acquired on the basis of the standard brain imaging protocol of the clinic, with section thickness and in-plane resolutions matching those of the synthetic MR imaging sequence. Total acquisition time was 11:00 minutes for all 4 2D conventional MR imaging sequences. Conventional 3D T1WI MPRAGE and 3D FLAIR sequences were additionally acquired (12:17 minutes) for gold standard radiologic and volumetric analyses. Synthetic PD-, T1-, and T2-weighted, and FLAIR images were obtained by using synthetic MR imaging (SyMRI) 7.2 RC software (Synthetic MR, Linköping, Sweden) with synthetic TE, TI, and TR matching the conventional images. To study the repeatability of the volumetric SyMRI measurements, we temporarily took each participant out of the scanner and repositioned them for a second synthetic MR imaging acquisition. Similarly, a second MPRAGE sequence was obtained after repositioning in 5 controls, to compare the repeatability of conventional and synthetic volumetrics.
Radiologic Assessment
All images were blindly and independently assessed by a neuroradiologist and a fellow in neuroradiology (F.H., C.C.) on standard radiologic workstations in the PACS in a random order in 2 sessions 2–4 weeks apart, in which each participant was only represented once per session with either the synthetic or conventional images. Image quality was rated on a 4-level scale: poor (artifacts or image-quality issues rendering the image quality insufficient for diagnostic purposes), sufficient (sufficient image quality for diagnostic purposes but noticeable artifacts or image-quality issues), good (only minor artifacts or image-quality issues), and excellent (no artifacts or image-quality issues). MS lesions were assessed by number and localization (periventricular, juxtacortical, infratentorial) and for the presence of black holes, defined as lesions with lower signal intensity than the normal-appearing white matter in T1WI (yes, no). Discrepancies were compared to the gold standard: the majority decision of the 2 original raters based on all available imaging data, including the 3D T1WI and FLAIR sequences, with any discrepancies resolved by a third rater, senior neuroradiologist (M.K.-W.). Due to the difficulty of quantifying and delineating confluent-versus-nonconfluent lesions for the lesion count, manual lesion segmentations were additionally performed on the 2D synthetic and conventional images by a resident in radiology (T.G.) in ITK-SNAP (www.itksnap.org).17
Contrast-to-noise ratio analyses were performed by manual tracing of 12 circular ROIs. Two were placed in the CSF (anterior horns of the lateral ventricles bilaterally), 6 in the GM (thalami, frontal cortex, occipital cortex bilaterally), and 4 in the WM (centrum semiovale bilaterally, genu and splenium), and the measurements were averaged for each tissue type. For patients with MS, an additional ROI was placed in the largest MS lesion. Care was taken to avoid regions with artifacts and tissue borders. The contrast-to-noise ratios were calculated by dividing the intensity difference of the tissues by the median SD of all 12 ROIs.
Volumetry
Volumetric measurements from synthetic MR imaging were obtained by using the automatic segmentation in SyMRI, Version 7.2 RC. Conventional volumetric measurements were obtained by using MPRAGE in FreeSurfer (http://surfer.nmr.mgh.harvard.edu), Statistical Parametric Mapping (SPM; http://www.fil.ion.ucl.ac.uk/spm/software/spm12), and the FMRIB Software Library (FSL; http://www.fmrib.ox.ac.uk/fsl).18⇓–20 Automatic lesion segmentation based on the conventional 3D images was performed in the Lesion Segmentation Toolbox (Technische Universität München, Munich, Germany) for SPM,21 and lesion filling and masking were performed in FSL and SPM to reduce the bias of MS lesions on GM segmentations.22 Segmentations were quality-controlled by a resident in radiology (T.G.), and segmentation parameters in FreeSurfer, FSL, SPM, and the Lesion Segmentation Toolbox were optimized to ensure accurate volumetric data. Initial FSL parameters were chosen on the basis of previous recommendations.23 Software versions, input, parameters, and processing times can be found in On-line Table 1.
The volumetric measurements of interest were the brain volume (BV), the white matter volume, the gray matter volume, and the intracranial volume (ICV). In the SyMRI software, lesion volume is not estimated, but there is a segmentation of the remaining intracranial content that is not recognized as WM, GM, or CSF. This segmentation class is henceforth referred to as “non-WM/GM/CSF” and contains not only MS lesions but also flow voids in larger blood vessels. The brain tissue measurements were further normalized to the ICV, resulting in the brain parenchymal fraction (BPF), white matter fraction (WMF), gray matter fraction (GMF), and non-WM/GM/CSF fraction. Examples of the synthetic segmentation output can be seen in Fig 1.
Statistical Analysis
Statistical analysis was performed in SPSS Statistics 23 (IBM, Armonk, New York). The Shapiro-Wilk normality test was used to investigate the normality of distribution. Group differences were studied by using the Wilcoxon signed rank test for ordinal dependent data and the McNemar test for binary dependent data. The Spearman rank correlation coefficient was used for correlation analysis. Repeatability was quantified by using the intermeasurement coefficient of variance (CoV) according to the following equation (A and B being the first and second measurements per patient): Statistical significance was an α level of 5% (2-tailed, equal variances not assumed).
Results
Image Quality and Diagnostic Accuracy
A representative example of conventional and synthetic MR images is seen in Fig 1. The diagnostic image quality and contrast-to-noise ratio measurements are given in Fig 2. Synthetic images were of sufficient or good quality overall, except for FLAIR images, which were degraded by basilar artery and CSF pulsation artifacts in 48% of the scans. Other image artifacts encountered in the synthetic MR images were chemical-shift displacement, discrete Gibbs ringing phenomenon along the superior sagittal sinus, and a sinusoidal intensity difference in the CSF in PD-weighted and T2WI in the anteroposterior direction, which was more easily detected if the lateral ventricles were enlarged. Artifacts are exemplified in On-line Fig 1. A proportion (23%) of the conventional spin-echo T1WIs were considered of poor diagnostic quality, mainly due to pulsation artifacts and low tissue contrast. Meanwhile, none of the synthetic T1WIs were rated of poor diagnostic quality, and synthetic T1WIs were assessed as having better image quality overall for diagnostic purposes (P < .001). The difference in tissue contrast was independently corroborated by objective differences in contrast-to-noise ratios, also illustrated in Fig 2.
All 40 participants were correctly classified as patients/controls with both MR imaging techniques, and incidental findings were identical: unspecific white matter changes in healthy controls (n = 3), pineal cysts (n = 3), mucosal thickening in the maxillary sinuses (n = 2), cerebrovascular lesions (n = 1), contusional injury (n = 1), choroid plexus cyst (n = 1), arachnoid cyst (n = 1), and abnormally shaped eye bulb (n = 1). The lesion count was higher in synthetic MR imaging compared with conventional MR imaging (median, 71 versus 64; interquartile range, 62 versus 56; P < .001), but this difference was on the same order of magnitude as the difference between the 2 raters (median, 74 versus 62; interquartile range, 68 versus 53), illustrated in Fig 3. Manual lesion segmentations corroborated this difference with higher manual lesion volumes in synthetic MR imaging (9.5 versus 9.3; interquartile range, 18 versus 16 mL; P = .009). The surplus in synthetic lesion volume was mainly due to inclusion of tissue considered “dirty-appearing” WM in the conventional images, exemplified by differences in manual lesion segmentations in On-line Fig 2. There was no statistical difference in the detection of periventricular lesions (P = .50, 1 false-negative finding on synthetic MR imaging), juxtacortical lesions (P = .50, 1 false-negative finding on synthetic MR imaging), or infratentorial lesions (P = .77, 1 false-negative finding on conventional MR imaging, 1 false-positive finding on synthetic MR imaging). There was no statistical difference in the detection of black holes (P = .50); 2 additional black holes were detected with synthetic MR imaging (1 false-negative finding on conventional MR imaging, 1 false-positive finding on synthetic MR imaging).
Repeatability and Feasibility of Volumetry
The SyMRI software provided volumetric measurements in the same order of magnitude as 3 commonly used volumetric software programs, illustrated in Fig 4. SyMRI was also the fastest segmentation method by several orders of magnitude (16 seconds compared with 12–187 minutes), without manual intervention, as reported in On-line Table 1. The test-retest repeatability of all volumetric methods is reported in Table 2. SyMRI had a lower repeatability error than FreeSurfer, FSL, and SPM for BV, ICV, BPF, and GMF. However, SyMRI had a similar error compared with the other volumetry methods regarding the segmentation of GM, WM, and WMF.
The SyMRI CoV for all 40 participants was 0.30% for BV and 0.23% for BPF, while the CoV was higher for WM and GM segmentations (gray matter volume and GMF, both 1.4%; white matter volume, 1.8%; WMF, 1.9%). The largest repeatability errors in the SyMRI segmentations of patients with MS were for the smallest tissue volumes and fractions: the non-GM/WM/CSF volume (CoV, 6.5%) and fraction (CoV, 6.1%). The average absolute and relative measurement differences for patients with MS and controls are reported in On-line Table 2. In terms of the clinical significance, the Expanded Disability Status Scale score was correlated with SyMRI BPF (r = −0.52, P = .02) and GMF (r = −0.53, P = .02), but not with WMF (r = −0.2, P = .42) or non-GM/WM/CSF fraction (r = 0.21, P = .37).
Discussion
This is the first implementation of synthetic MR imaging on a Siemens MR imaging system and, to the best of our knowledge, the first independent evaluation of the diagnostic accuracy of the technique in MS. In this initial study, we found that synthetic MR imaging may be feasible as an alternative or complement to conventional PD-, T1-, and T2WI in MS. Differences in the contrast of lesions and dirty-appearing WM caused discrepancies in the lesion count and volume, but these differences did not affect the diagnostic classification or lesion topography. However, pulsation artifacts in the synthetic FLAIR imaging must be addressed before it can substitute for conventional FLAIR. Similar FLAIR artifacts have previously been reported in an implementation on a 1.5T scanner from another manufacturer.11 Artifacts in the synthetic images could also be seen in the preprocessed DICOM files and were therefore not related to a postprocessing issue in the image reconstruction of the SyMRI software. Implementation of the sequence on the newer Siemens E11 software platform and on 1.5T scanners is underway, and we hope to re-evaluate these issues in the next implementation. Although incidental findings were identical with the 2 MR imaging techniques, future studies will have to evaluate the diagnostic accuracy of synthetic MR imaging in pathologies other than MS. The time-savings compared with conventional MR imaging was 7% for PD-, T1-, and T2WI (6:50 versus 7:22 minutes).
We have further shown that the volumetric measurements in the SyMRI are in agreement with those in other volumetric software and that SyMRI had the lowest repeat measurement errors for BV, ICV, BPF, and GMF among all tested volumetric methods. This finding is likely because the sequence inherently has information about PD, the longitudinal and transverse relaxation rates, and B1 field inhomogeneities, which can render more precise delineations of the intracranial and brain surfaces compared with just using information from T1WI. The multiparametric acquisition is equivalent to a multichannel approach, but without the need for image coregistration because the synthetic volumetrics are based on a single sequence. Because patients with MS are followed longitudinally, normalized brain volumetrics are of special interest. BPF was measured in patients with MS with SyMRI with repeatability errors (CoV, 0.25%; average measurement difference, 0.06%) well below the reported annual brain atrophy rate (0.5%–1.3%).24⇓⇓–27 The variability in measurements was larger in WMF and GMF (CoV, 1.2% and 1.7%; average measurement difference, 0.4% and 0.5%), but still lower than or equal to reported atrophy rates. The segmentation of the non-GM/WM/CSF fraction, which is the smallest tissue compartment, had the largest test-retest variability (CoV, 6.1%; average measurement difference, 1.6%) but may be of clinical importance because it is expected to partly reflect the lesion burden in MS. For comparison, MS lesion volume increases by approximately 5%–10% annually in MS.28,29 These results imply that GM, WM, and non-GM/WM/CSF measurements should preferably be studied in perspectives longer than 1 year. From a clinical perspective, the SyMRI was the most practical segmentation method because it could be launched in the PACS and delivered robust volumetrics within a few tens of seconds. In radiologic practice, the feasibility of SyMRI volumetry lowers the threshold for being able to deliver quantitative biomarkers, which is becoming increasingly requested by our clinical colleagues.
This study has several strengths, such as the evaluation of both the diagnostic and volumetric output of SyMRI. The consecutive recruitment of the patients with MS is also advantageous because it reflects a typical panorama of patients with MS in a neurologic outpatient clinic, with all 3 subtypes of MS represented with a wide range of disease duration and Expanded Disability Status Scale scores, as well as the use of a control group. Limitations include a relatively small sample size and that repeat MPRAGE sequences could only be obtained in a subset of controls due to the length of the combined imaging protocol with reproducibility scans. In this study, we primarily harmonized settings in terms of voxel size and timing parameters, but unfortunately, the receiver bandwidth was not optimized in the same way. A lower bandwidth results in a higher signal-to-noise ratio and therefore hampers this comparison, but it also means that SyMRI was affected with chemical-shift displacement artifacts and theoretically longer minimum TEs and TRs, longer echo spacing, and increased susceptibility artifacts.
In terms of generalizability, the study is limited by the use of a 4-mm section thickness with a 30% gap, while a 3-mm section thickness is recommended for MS studies with 2D sequences.30 Subsequently, we have been able to reduce this section thickness to 3 mm, but the gap remains to reduce cross-talk between sections. Future technical effort should try to eliminate the need for this gap in SyMRI and should try to make high-resolution 3D acquisitions possible, which is increasingly important in MS studies.30,31 Previous studies of the effects of intravenous gadolinium-based contrast media on the volumetric output of SyMRI indicate that it is important to consistently acquire synthetic MR imaging either pre- or postcontrast for comparability32; these findings are the reason we chose to focus on nonenhanced acquisitions in this study. Future studies will be able to clarify how synthetic MR imaging could be implemented as a postcontrast examination and in a longitudinal fashion in MS. Although the basic concepts of synthetic MR imaging implementation on other MR imaging systems are similar, a comparative study across 1.5 and 3T scanners from the 3 major MR imaging systems is warranted. Lastly, an advantage with SyMRI that remains to be studied is the possibility of being able to adjust the synthetic TR, TE, and TI parameters post hoc, which may potentially allow the user to optimize the contrast between normal and pathologic tissues.
Conclusions
This study shows that synthetic MR imaging can be implemented on Siemens MR imaging scanners and can be an alternative to conventional MR imaging for generating diagnostic PD-, T1-, and T2-weighted images, but not yet FLAIR images, in patients with MS and healthy controls. Volumetric segmentations can be obtained with a few tens of seconds of processing without any extra scanning time in SyMRI. The measurements of BV and BPF have an excellent repeatability and are thus feasible for longitudinal studies in MS.
Acknowledgments
We would like to thank the participants of the study and the staff at the Karolinska University Hospital MR imaging unit in Huddinge. We also acknowledge Stina Granberg, who has been helpful with graphic assistance. We further thank Karin Johnsson, Eric Westman, and Russell Ouelette for helpful comments on this article.
Footnotes
Disclosures: Tobias Granberg, Farouk Hashim, Yngve Forslin, Carmen Cananau, Sara Shams, Peter Aspelin, Maria Kristoffersen-Wiberg—RELATED: Grant: ALF Grant from Karolinska Institutet and Stockholm City Council.* Sten Fredrikson—UNRELATED: Payment for Lectures (including service on Speakers Bureaus): Allergan, Bayer, Biogen Idec, Genzyme, Merck, Sanofi, Teva; Payment for Development of Educational Presentations: Teva. *Money paid to the institution.
This work was supported by Karolinska Institutet and Stockholm County Council through an ALF grant.
Paper previously presented at: Annual Meeting of the Radiological Society of North America, November 29–December 1, 2015; Chicago, Illinois.