Abstract
Summary: A glaucoma drainage implant was detected on plain skull radiographs before MR imaging examination of the brain. The patient was denied the MR imaging for fear of dislodging the “metallic foreign body.” The Baerveldt glaucoma drainage implant was mistakenly identified as an orbital metallic object based on its radiographic characteristics. Because none of the current glaucoma drainage implants contain ferromagnetic material, patients with these devices can undergo MR imaging without special precaution.
Drainage implants are increasingly being used for the surgical treatment of glaucoma refractory to medical and standard surgical therapy. Echographic and radiographic findings may suggest the diagnosis of an orbital foreign body if the ophthalmic history is not known.
Case Report
An asymptomatic 82-year-old female patient with open angle glaucoma presented to her ophthalmologist for evaluation of a “metallic object” in her left orbit (Fig 1). The object in question was identified as a 350-mm2 Baerveldt glaucoma drainage implant (Pharmacia Co., Kalamazoo, MI) that had been previously placed to lower intraocular pressure. The smooth convex radiographic profile of the device corresponded to the globe curvature in the superotemporal orbit.
Discussion
Of the commercially available glaucoma drainage implants, only the Baerveldt glaucoma drainage implant (250, 350, and 425 mm2) is radiopaque (Fig 2). Barium-impregnated silicone used in the fabrication of the Baerveldt plate increases echographic resolution and facilitates radiographic identification. The Krupin-Denver eye valve to disc implant (184 mm2) (E. Benson Hood Laboratories, Pembroke, MA), Ahmed glaucoma valve (184 mm2) (New World Medical, Rancho Cucamonga, CA), Molteno drainage device (single plate 129 mm2 and double plate 259 mm2) (Molteno Ophthalmic Ltd., Dunedin, New Zealand), and the Joseph valve (Valve Implants Limited, Hertford, England) feature a translucent silicone tube connected to a plate made of rigid plastic or flexible silicone rubber that is fixed to the sclera (Fig 3). None of these devices contain ferromagnetic material or have been impregnated with a radiopaque substance.
Intraocular pressure is lowered when aqueous humor flows from inside the eye through the tube into the space between the plate that rests on the scleral surface and surrounding fibrous capsule. The adjacent periocular tissues absorb aqueous humor from the reservoir. Although intraocular ferromagnetic foreign bodies have been shown to move during MR imaging (1–3), because neither the Baerveldt drainage implant nor any of the other commercially available devices contain ferromagnetic material, inadvertent dislocation of these devices is unlikely.