Abstract
BACKGROUND AND PURPOSE: Recent improvements in MR gradient technology allow significant increases in diffusion weighting without prohibitive signal-to-noise degradation. The purpose of our investigation was to establish normative references for the signal intensity characteristics and apparent diffusion coefficient values of the adult brain at high b values.
METHODS: Fifty adults underwent diffusion-weighted single-shot spin-echo echo-planar MR imaging. Isotropic diffusion-weighted images were obtained with b values of 0, 1000, 2000, 2500, 3000, and 3500 s/mm2. Qualitative assessments were made in multiple regions of interest in gray and white matter. Three apparent diffusion coefficient maps were generated for each of six patients with a 2-point technique at a b value of 0 and at b values of 1000, 2000, and 3000 s/mm2.
RESULTS: Increasing b values result in a progressive decrease in the gray to white matter signal intensity ratio. Isointensity between gray and white matter results at b values between 1000 and 2000 s/mm2. At b values greater than 2000, the gray-white pattern reverses relative to the usual b value of 1000. Apparent diffusion coefficient values were shown to decrease with increasing b values.
CONCLUSION: Attention to the reversal of gray-white contrast and the dependence of apparent diffusion coefficient on the b value are important in avoiding erroneous assignment of pathologic abnormalities to normal regions. This study provides the normative data for future diffusion investigations performed at high b values.
Recent improvements in MR gradient technology allow significant increases in diffusion sensitivity without prohibitive signal-to-noise degradation. Understanding the appearance of images obtained with higher b values is critical for the assessment of future clinical applications and investigations using this technology. The ability of current diffusion-weighted imaging to detect and delineate restricted diffusion is well documented (1–5). The purpose of our investigation was to establish normative standards for the signal intensity characteristics and trace apparent diffusion coefficient (ADC) values of the adult brain at varying high b values.
Increased gradient amplitudes permit higher b values and greater diffusion sensitivity. Gradient subsystems available on clinical imagers with echo-planar imaging have been limited to diffusion-weighted imaging with b values up to 1200 s/mm2. Newer clinical imagers can provide b values of 3500 s/mm2 with echo times less than 100 ms. Isotropic diffusion-weighted images at the standard diffusion gradient strength of 1000 s/mm2 reveal relative hyperintensity of cortical gray matter compared with white matter. The signal characteristics of adult brain at higher b values have not been reported.
This investigation aimed to establish normative references for the signal intensity characteristics and trace ADC measurements of the adult brain with diffusion-weighted spin-echo echo-planar imaging at varying high b-value diffusion sensitivities. Furthermore, this study provides the normative data for future diffusion investigations performed at high b values.
Methods
Participants
Fifty healthy adults, ranging in age from 17 to 65 years, underwent diffusion-weighted MR imaging. Patients were referred with complaints of headache and for screening for possible neoplasm or multiple sclerosis. Some served as control participants for various concurrent functional imaging studies. Participants were excluded if underlying pathologic abnormalities were revealed by standard imaging sequences.
Imaging
All images were obtained with a 1.5-T clinical imager optimized for neuroimaging with a 150-T/m/s slew rate and a 40-mT/m maximum gradient strength. Single-shot spin-echo echo-planar imaging isotropic diffusion-weighted images were produced with b values of 0, 1000, 2000, 2500, 3000, and 3500 s/mm2. Other imaging parameters included 10000/72–96 (TR/TE); field of view, 38 × 19 cm; matrix, 192 × 128; section thickness, 5 mm; section gap, 0 mm; and number of acquisitions, one to four. Higher numbers of acquisitions and minimum echo times were used to improve signal-to-noise ratios at higher b values in 43 of 50 participants. The number of acquisitions was adjusted for each b value: b = 1000 images, number of acquisitions = one; b = 2000 images, number of acquisitions = two; b = 3000 images, number of acquisitions = four.
Quantitative assessment of trace ADC measurements and of signal intensity variation with b-value modulation was conducted for six of the 50 participants using a fixed echo time and variable numbers of excitation. One of the 50 participants was studied with a fixed number of excitations to control for potential influence on the ADC measurement while the echo time was held constant. This participant also underwent imaging with a fixed number of excitations and variable echo times of 96 and 72 ms at a b value of 1000 to assess the effect of echo time modulation on ADC measurement. The ADC data regarding this participant were not averaged with those of the other six because of the differing numbers of excitation.
The signal characteristics of diffusion-weighted imaging sequences were qualitatively assessed for all 50 participants and are reported together because they were independent of the number of signal averages and echo time. In the spin-echo echo-planar diffusion-weighted imaging pulse sequence, the duration of diffusion bipolar gradient pulses and the duration between the two bipolar gradient pulses were held constant for different b values; hence, different b values were obtained by changing the gradient strength of the bipolar gradient pulses while keeping the diffusion time constant.
Data Analysis
Qualitative assessments were made of gray-white contrast and regions of interest (ROI) in the cortical gray matter, centrum semiovale, caudate nuclei, lentiform nuclei, thalami, internal capsules, central pons, middle cerebellar peduncles, and dentate nuclei for isotropic diffusion-weighted images. Quantitative ADC measurements were made from ADC images for the same ROI for six participants.
The diffusion-weighted imaging signal intensity ratio of putamen to centrum semiovale white matter was calculated for b values of 1000, 2000, and 3000 s/mm2 for the six participants evaluated with a fixed echo time and variable numbers of excitation. Although the signal intensity of putamen is not always isointense to cortical gray matter, the putamen was chosen to avoid the issue of partial volume averaging of CSF in the measurement of gray matter signal intensity. The centrum semiovale is heterogeneous on diffusion-weighted imaging, particularly at higher b values. Therefore, measurements were independently obtained for two areas of white matter that were visually assessed to have the greatest and least signal intensity on the diffusion-weighted images. All measurements were obtained using a developmental software program on a separate workstation.
Image Processing
The isotropic diffusion-weighted images were generated using the cube root of the products of the signal intensities obtained from images sensitized for diffusion along each of the three orthogonal axes (ie, (IxIyIz)1/3). Quantitative ADC measurements were made from ADC images created with a two-point linear fit approach. This method involves generating images with two different b values (b0 = 0 and b1 = X s/mm2) and then solving the equation for the diffusion coefficient. For the two-point case: where ln is the natural logarithm, and S0 and S1 are the image intensities obtained with b values of b0 and b1, respectively. ROI size was 55 mm2 for all measurements. CSF spaces were avoided to minimize partial volume averaging.
Results
Gray-White Contrast
The pattern of gray-white matter contrast observed on the diffusion-weighted imaging obtained with b values of 1000 s/mm2 is similar to the contrast features of T2-weighted images, with gray matter hyperintense to white matter. Increasing b values result in progressively greater diffusion weighting while diminishing the appearance of T2 weighting. At b values greater than 2000, the gray-white pattern reverses relative to what is observed at the usual b value of 1000. The most pronounced reversal occurs at a b value of 3500 s/mm2 (Fig 1). Isointensity between cortical gray and white matter accordingly results at b values between 1000 and 2000 s/mm2, although the precise value corresponding to isointensity was not determined.
ROI Assessment
Several regions of normal brain are relatively hyperintense on diffusion-weighted imaging performed at lower diffusion gradient strengths. These include the posterior limbs of the internal capsules, the mesial frontal lobe cortex, the central midbrain and pons, and the insular cortex. The degree of hyperintensity in the mesial frontal lobe cortex and the insular cortex was greatest on the diffusion-weighted images obtained with a b value of 1000, gradually becoming isointense with surrounding cortex on images obtained with progressively higher b values (Fig 1). The focal assessment of ROI in subcortical white matter structures revealed the general trends toward hyperintensity on the images obtained with b values greater than 2000 s/mm2. This was most evident in the subcortical white matter, centrum semiovale, posterior internal capsule, and central brain stem regions exhibiting relative hyperintensity. These signal characteristics were independent of echo time and number of excitations.
Signal intensity ratios provided a quantitative assessment of the gray-white matter contrast on diffusion-weighted images (Table 1). These measurements indicate a decline in the gray-white matter signal intensity ratio with increasing b values for ROI in the putamen and centrum semiovale white matter.
ADC Measurements
ADC values decreased with increasing diffusion weighting. Greater SDs were noted for ROI adjacent to CSF spaces and for ADC maps derived from the single number of excitations with a b value of 1000 (Fig 2 and Table 2). ADC measurements were shown to vary with diffusion sensitivity, independent of number of excitations (Table 3). Echo time did not affect the mean ADC values measured at a fixed number of excitations (Table 4). Differentiation of gray and white matter on the ADC maps improved with increasing b values (Fig 3).
Discussion
Diffusion imaging at standard b values is a remarkably useful tool in the detection and delineation of restricted diffusion in the clinical setting, particularly in the evaluation of acute ischemia. Understanding the appearance of images obtained at standard and higher b values is critical for the assessment of future clinical applications and investigations using this technology. Our investigation provides normative references for the signal intensity characteristics and ADC values of the adult brain at varying high b-value diffusion sensitivities. The qualitative and quantitative features of diffusion images obtained with varying b values provide an opportunity to gain insight into the physiological basis of diffusion imaging as different characteristics of tissue are revealed at different diffusion sensitivities. Increasing the acquisition b value has several implications for diffusion-weighted imaging. The gray-white contrast reversal results in an alteration of the lesion-to-background contrast. The relative conspicuity of restricted diffusion in gray matter is expected to be greater at higher b values, whereas lesion conspicuity in white matter may be greater at lower b values. The optimal b value for evaluation of specific disease states with diffusion-weighted imaging must be determined. The use of multiple sequences with differing b values may yield complementary results (Fig 4).
The pattern of gray-white matter contrast in the diffusion-weighted images obtained with b values of 1000 s/mm2 is likely a result of the residual T2 weighting. The artifactual hyperintensity or “T2 shine-through” seen on diffusion-weighted images of lesions with T2 hyperintensity is also likely a result of the residual T2 weighting present in the diffusion trace images. Higher b-value images have relatively less intrinsic T2 weighting and may be useful in differentiating restricted diffusion from T2 shine-through (Fig 4).
The changes in the relative diffusion-weighted imaging signal intensity of gray and white matter with b-value modulation are predicted by the signal intensity equation for a spin-echo diffusion-weighted image, which is given by: assuming monoexponential diffusion, where N(h) is the proton density, TR is the pulse repetition time, T1 is the spin-lattice relaxation time, TE is the spin-echo time, T2 is the spin-spin relaxation time, b is the strength of the diffusion gradient, and D is the diffusion coefficient. Using this equation, we plotted the theoretical signal intensities for gray and white matter by using average imaging parameters and tissue relaxation values (6, 7). From these data, it is possible to derive hypothetical signal intensity curves for a range of b values (Fig 5). The progressive increase in white matter intensity and decrease in gray matter intensity on diffusion-weighted images with increasing b values shown empirically by our data (Fig 1 and Table 1) are predicted by this model.
Table 2 unambiguously shows a consistent pattern of decreasing ADC measurements for increasing b values. For some ROI, the ADC decreased approximately 30% to 35% between b values of 1000 and 3000 s/mm2. These data indicate that the monoexponential assumption underlying these calculations is inappropriate. If this assumption were valid, as the b values increased, the ADC would remain constant for all calculations. A likely explanation of these data is that the apparent diffusion is more accurately represented by a biexponential response. Rapidly decaying and slowly decaying diffusion components have been described in both animal and human models (8–10). At relatively low b values, signal intensity in the trace images is dominated by a fast diffusion component, Dfast, whereas at higher b values, signal intensity is contributed largely by the slower component of diffusion, Dslow. Consequently, as b values are increased from 1000 to 3000 s/mm2, the signal intensity in the trace images transitions from a Dfast to a Dslow weighting. There are also important differences between intracellular (Din) and extracellular (Dex) diffusion coefficients corresponding to the slow and fast diffusion components, respectively. Niendorf et al (9) showed that the Dfast and Dslow do not equal but are proportional to Din and Dex, respectively. Kraemer et al (10) showed that the extracellular component dominates at b values less than 1000s/mm2 and that the ADC measurements obtained at these values primarily reflect extracellular diffusion, especially when the data are fitted to a single exponential.
A multi-component diffusion model has substantial implications for the postprocessing of diffusion trace (ADC) images. Obviously, the two-point calculation is inadequate to characterize the signal intensity as the b value is modulated. The least-squares multi-point approach, although possibly reducing the SD of the measurements, will yield incorrect results. A more sophisticated postprocessing approach is required for these data that uses the biexponential nature of the signal attenuation. One processing avenue to pursue is the non-negative least-squares algorithm that theoretically would extract all diffusion components. However, it is likely that this would require increased b-value sampling density as well as higher b values to improve the resolution of the diffusion measurements. At the very least, this technique would significantly increase the acquisition time necessary to generate the data sets for these calculations.
Burdette et al (11) compared ADC measurements using two-point and six-point techniques in the b-value range of 0 to 1000 s/mm2 and noted that the mean percentage difference between the two methods was 0.84%. Their results support the concept that the ADC calculation for a two-point calculation is practical, reproducible, and acceptable for clinical diffusion examinations. We observed significant variation in the ADC value for b values of 1000 to 3000 s/mm2, well beyond the range that was investigated by Burdette et al. For practical purposes, our results suggest that it is not valid to extrapolate ADC values obtained at a particular b value to other b values.
The large SDs of the ADC measurements, particularly at ROI near CSF spaces, result in some overlap in ADC values for normal gray and white matter structures. This may impact the quantitative values of the ADC measurements in routine clinical imaging. The decrease in SDs with increasing b values for the six participants averaged in Table 2 was likely a result of signal averaging. The SDs for the participant for whom a fixed number of excitations was used were not b-value-dependent (Table 3). Increasing numbers of excitation and careful ROI placement may reduce the SDs of ADC measurements. ADC measurement was not affected by echo time modulation in the single participant evaluated (Table 4).
Ulug et al (12) addressed the problem of ADC reproducibility within and between neuroanatomic sites. The relatively large SDs of ADC measurements observed in this study again emphasize this issue. The results presented by Ulug et al showed minimal interparticipant variation when comparing ipsilateral to contralateral ratios of ADC as a means to quantitate changes in ADC measurements.
The usefulness of ADC maps has been questioned by other investigators who cite a greater sensitivity and accuracy and a very high specificity in the detection of early cerebral infarctions (98.1, 97.7, and 97.1%, respectively) for orthogonal axis diffusion-weighted images compared with ADC maps (13, 14). The validation of their conclusions in the evaluation of stroke at higher b values is necessary. The need for postprocessing, specifically ADC map generation, may be less at higher b values as the intrinsic diffusion sensitivity is increased and the T2 effects are reduced.
The variation in the echo time and the number of signal averages between and within the various subgroups of participants in this study could be considered a limitation of this investigation. The inclusion of participant data obtained with these variations has both positive and negative effects. The qualitative observations regarding signal intensity features were correct for all patients independent of these variations, thus increasing the likelihood of reproducibility by other centers. Conversely, the variations in protocol limited the number of examinations available for quantitative assessment of ADC variation independent of echo time and number of excitations. Measurements were obtained from six participants with a fixed echo time. Averaging of a larger number of participants' data may provide a more reliable standard of reference for normal ADC values.
Conclusion
Interpretation of diffusion-weighted images obtained with b values greater than 1000 s/mm2 must consider the increasing contribution of diffusion weighting and the diminished T2 effects. Diffusion-weighted images obtained at b values greater than 2000 s/mm2 have contrast features more accurately reflecting the relative ADC values of gray and white matter, with hyperintensity corresponding to lower ADC values. Restricted diffusion and lower ADC values in white matter results in hyperintensity relative to gray matter on the more heavily diffusion-weighted images. ADC dependence on b value must be considered in the interpretation of these quantitative measurements. Our data indicate that diffusion in human brain is not monoexponential. Although ADC algorithms using the monoexponential assumption are practical, a more complex approach is required to extract the fast and slow diffusion components present in brain tissue. Attention to these findings, especially the reversal of gray-white contrast and ADC dependence on b value, is important in avoiding erroneous assignment of pathologic abnormalities to normal regions. Future studies are necessary to define the role of high b-value diffusion imaging further in the evaluation of acute stroke and other pathologic abnormalities.